Zno ZnO ZnCO3 ZnS Zn(SO4)    ZnSe   ZnTe ZnF2  ZnCu     Basic 

XPS Spectra
Zinc (Zn) Compounds
The XPS Spectra section provides raw and processed survey spectra, chemical state spectra, BE values, FWHM values, and overlays of key spectra.
Atom% values from surveys are based on sample, as received, and Scofield cross-sections. Atom% values are corrected for IMFP and PE.
Peak-fits are guides for practical, real-world applications. Peak-fits are not fully optimized or designed to test any theory.


Zinc Carbonate  (Zn(CO3) – Smithsonite)
Survey, Peak-fits, BEs, FWHMs, and Peak Labels


 Periodic Table   → Six (6) BE Tables
Survey Spectrum from Zn(CO3)
Freshly exposed bulk, Flood gun is ON, C (1s) BE = 285.0 eV, Ag (3d5/2) FWHM = 1.3 eV

 Periodic Table  → Six (6) BE Tables
Zn (2p) Spectrum from Zn(CO3) Raw
Fresh exposed bulk, Flood gun is ON, C (1s) BE = 285.0 eV, Ag FWHM = 0.75 eV
Zn (2p) Spectrum from Zn(CO3) – Peak-Fit
Freshly exposed bulk, Flood gun is ON, C (1s) BE = 285.0 eV, Ag FWHM = 0.75 eV

 


 
Zn (2p) Spectrum from Zn(CO3) – Extended Range
Fresh exposed bulk, Flood gun is ON, C (1s) BE = 285.0 eV, Ag FWHM = 0.75 eV
Zn (2p) Spectrum from Zn(CO3) – Raw – Vertically Expanded
Freshly exposed bulk, Flood gun is ON, C (1s) BE = 285.0 eV, Ag FWHM = 0.75 eV

 Periodic Table  → Six (6) BE Tables
O (1s) Spectrum from Zn(CO3) – Raw
Freshly exposed bulk, Flood gun is ON, C (1s) BE = 285.0 eV, Ag FWHM = 0.75 eV
O (1s) Spectrum from Zn(CO3) – Peak-fit
Freshly exposed bulk, Flood gun is ON, C (1s) BE = 285.0 eV, Ag FWHM = 0.75 eV

 Periodic Table  → Six (6) BE Tables
C (1s) Spectrum from Zn(CO3) – Peak-fit – As measured 
Freshly exposed bulk, Flood gun is ON, C (1s) BE = 284.41 eV, Ag FWHM = 0.75 eV
C (1s) Spectrum from Zn(CO3) – Peak-Fit – Corrected
Freshly exposed bulk, Flood gun is ON, C (1s) BE = 285.0 eV, Ag FWHM = 0.75 eV

 Periodic Table  → Six (6) BE Tables
Zn (3s) Spectrum from Zn(CO3) – Raw
Freshly exposed bulk, Flood gun is ON, C (1s) BE = 285.0 eV, Ag FWHM = 0.75 eV
Zn (3s) Spectrum from Zn(CO3) Peak-Fit
Freshly exposed bulk, Flood gun is ON, C (1s) BE = 285.0 eV, Ag FWHM = 0.75 eV
   


 Periodic Table  → Six (6) BE Tables
Zn (3p) Spectrum from Zn(CO3) – Raw
Freshly exposed bulk, Flood gun is ON, C (1s) BE = 285.0 eV, Ag FWHM = 0.75 eV
Zn (3p) Spectrum from Zn(CO3) Peak-Fit
Freshly exposed bulk, Flood gun is ON, C (1s) BE = 285.0 eV, Ag FWHM = 0.75 eV

 


 Periodic Table  → Six (6) BE Tables
  Zn (KLL) Auger Spectrum from Zn(CO3) – Raw
Freshly exposed bulk, Flood gun is ON, C (1s) BE = 285.0 eV, Ag FWHM = 0.75 eV
Valence Band Signals from Zn(CO3) – Raw
Freshly exposed bulk, Flood gun is ON, C (1s) BE = 285.0 eV, Ag FWHM = 0.75 eV

 
Overlays
 Periodic Table  → Six (6) BE Tables
Valence Band SpectraOverlay of Zno and Zn(CO3)
Freshly exposed bulk, Flood gun is ON, C (1s) BE = 285.0 eV, Ag FWHM = 0.75 eV
Zn (2p) SpectraOverlay of Zno and Zn(CO3)
Freshly exposed bulk, Flood gun is ON, C (1s) BE = 285.0 eV, Ag FWHM = 0.75 eV


 Periodic Table  → Six (6) BE Tables
Zn (2p) Spectra – Overlay of ZnO and Zn(CO3)
Freshly exposed bulk, Flood gun is ON, C (1s) BE = 285.0 eV, Ag FWHM = 0.75 eV
Zn Spectra – Overlay of Zno plasmons and Zn(CO3) E losses
Freshly exposed bulk, Flood gun is ON, C (1s) BE = 285.0 eV, Ag FWHM = 0.75 eV


Zn (2p) Spectra – Overlay of ZnO and Zn(CO3) losses – Expanded
Freshly exposed bulk, Flood gun is ON, C (1s) BE = 285.0 eV, Ag FWHM = 0.75 eV
Zn (2p3/2) Spectra – Overlay of ZnO and Zn(CO3)
Freshly exposed bulk, Flood gun is ON, C (1s) BE = 285.0 eV, Ag FWHM = 0.75 eV


End-of-spectra

Price to purchase raw data sets:
Raw spectra – VAMAS ASCII format ($6)
Raw spectra – SDP binary format ($5)
SDP v9 – $145 (3 yr license)

 



 

 

Transmission Function Tests


 

December 2015 – Transmission Function of Thermo K-Znpha Plus

 Periodic Table 
Survey Spectra of Ion Etched Copper (Cr), PEs = 50, 100, 150 and 200 eV

 Periodic Table 
March 2016 – Transmission Function of Thermo K-Znpha Plus 
 
Survey Spectra of Ion Etched Copper (Cr), PEs = 100, 150 and 200 eV

 Periodic Table  
August 2019 – Transmission Function of Thermo K-Znpha Plus
 
Survey Spectra of HOPG (C), PEs = 20, 50, 100 and 200 eV

 Periodic Table  
January 2022 – Transmission Function of Thermo K-Znpha Plus
Survey Spectra of Ion Etched Copper (Cr), PEs = 100, 120, 140, 160, 180 and 200 eV


 

Six (6) Chemical State Tables of Zn (2p) BEs

 

  • The XPS Library Spectra-Base
  • PHI Handbook
  • Thermo-Scientific Website
  • XPSfitting Website
  • Techdb Website
  • NIST Website

→  Periodic Table 

 



 

Notes of Caution when using Published BEs and BE Tables from Insulators and Conductors:

  • Accuracy of Published BEs
    • The accuracy depends on the calibration BEs used to calibrate the energy scale of the instrument.  Cu (2p3/2) BE can vary from 932.2 to 932.8 eV for old publications
    • Different authors use different BEs for the C (1s) BE of the hydrocarbons found in adventitious carbon that appears on all materials and samples.  From 284.2 to 285.3 eV
    • The accuracy depends on when the authors last checked or adjusted their energy scale to produce the expected calibration BEs
  • There are uncertainties and error ranges in nearly all BEs 
    • Flood guns
  • Worldwide Differences in Energy Scale Calibrations
    • For various reasons authors still use older energy scale calibrations
    • Some authors still adjust their energy scale so Cu (2p3/2) appears at 932.2 eV or 932.8 eV because this is what the maker taught them
    • This range causes BEs in the higher BE end to be larger than expected
    • This variation increases significantly above 600 eV BE
  • Charge Compensation
    • Samples that behave as true insulators normally require the use of a charge neutralizer (electron flood gun with or without Ar+ ions) so that the measured chemical state spectra can be produced without peak-shape distortions or sloping tails on the low BE side of the peak envelop.
    • Floating all samples (conductive, semi-conductive, and non-conductive) and always using the electron flood gun is considered to produce more reliable BEs and is recommended.
  • Charge Referencing Methods for Insulators
    • Charge referencing is a common method, but it can produce results that are less reliable.
    • When an electron flood gun is used, the BE scale will usually shift to lower BE values by 0.01 to 5.0 eV depending on your voltage setting. Normally, to correct for this flood gun induced shift, the BE of the hydrocarbon C (1s) peak maximum from adventitious carbon is used to correct for the charge induced shift.
    • The hydrocarbon peak is normally the largest peak at the lowest BE.
    • Depending on your preference or training, the C (1s) BE assigned to this hydrocarbon peak varies from 284.8 to 285.0 eV.  Other BEs can be as low as 284.2 eV or as high as 285.3 eV
    • Native oxides that still show the pure metal can suffer differential charging that causes the C (1s) and the O (1s) and the Metal Oxide BE to be larger
    • When using the electron flood gun, the instrument operator should adjust the voltage and the XY position of the electron flood gun to produce peaks from a strong XPS signal (eg O (1s) or C (1s) having the most narrow FWHM and the lowest experimentally measured BE.

→  Periodic Table 



Table #1

Zn (2p) Chemical State BEs from:  The XPS Library Spectra-Base”

C (1s) BE = 285.0 eV for TXL BEs
and C (1s) BE = 284.8 eV for NIST BEs

Element Atomic # Compound As-Measured by TXL or NIST Average BE Largest BE Hydrocarbon C (1s) BE  Source
Zn 12 Zn(OH)2 (N*1) 49.5 eV 284.8 Avg BE – NIST
Zn 12 Zn – element 49.7 eV 285.0 The XPS Library
Zn 12 Zn-O (N*3) 50.3 eV 51.1 eV 284.8 Avg BE – NIST
Zn 12 Zn-Si 50.4 eV 285.0 The XPS Library
Zn 12 ZnF2 (N*2) 50.9 eV 51.0 eV 284.8 Avg BE – NIST
Zn 12 Zn-(OH)2 51.3 eV 285.0 The XPS Library
Zn 12 ZnO native 51.4 eV 286.8 The XPS Library
Zn 12 ZnF2 52.0 eV 285.0 The XPS Library
Zn 12 Zn-O 52.0 eV 285.0 The XPS Library
Zn 12 Zn-CO3 52.0 eV 285.0 The XPS Library

 

Charge Referencing

  • (N*number) identifies the number of NIST BEs that were averaged to produce the BE in the middle column.
  • Binding Energy Scale Calibration expects Cu (2p3/2) BE = 932.62 eV and Au (4f7/2) BE = 83.98 eV.  BE (eV) Uncertainty Range:  +/- 0.2 eV
  • Charge Referencing of insulators is defined such that the Adventitious Hydrocarbon C (SO) BE (eV) = 285.0 eV.  NIST uses C (1s) BE = 284.8 eV 
  • Note:   Ion etching removes adventitious carbon, implants Ar (+), changes conductivity of surface, and degrades chemistry of various chemical states.
  • Note:  Ion Etching changes BE of C (1s) hydrocarbon peak.
  • TXL – abbreviation for: “The XPS Library” (https://xpslibrary.com).  NIST:  National Institute for Science and Technology (in USA)

 →  Periodic Table 



Table #2

Zn (2p) Chemical State BEs from:  “PHI Handbook”

C (1s) BE = 284.8 eV

Copyright ©:  Ulvac-PHI



Table #3

Zn (2p) Chemical State BEs from:  Thermo-Scientific” Website

C (1s) BE = 284.8 eV

Chemical state Binding energy
Zn (1s)
Zn 1303 eV
Zn native oxide 1304 eV
ZnCO3 1305 eV

 Periodic Table 

Copyright ©:  Thermo Scientific website



Table #4

Zn (2p) Chemical State BEs from:  “XPSfitting” Website

Chemical State BE Table derived by Averaging BEs in the NIST XPS database of BEs
C (1s) BE = 284.8 eV

 

Copyright ©:  Mark Beisinger



Table #5

Zn (2p) Chemical State BEs from:  “Techdb.podzone.net” Website

XPS Spectra – Chemical Shift | Binding Energy
C (1s) BE = 284.6 eV

XPS(X線光電子分光法)スペクトル 化学状態 化学シフト ケミカルシフト

Element Level Compound B.E.(eV) min max
Zn 2p Zn(OH)2 49.5 ±0.3 49.2 49.8
Zn 2p Zn 49.7 ±0.2 49.5 49.8
Zn 2p Zn2Cu 49.9 ±0.3 49.6 50.1
Zn 2p ZnAl2O4 50.4 ±0.3 50.1 50.6
Zn 2p Zn3Bi2 50.7 ±0.2 50.5 50.9
Zn 2p ZnF2 51.0 ±0.2 50.8 51.2

 

Histograms of NIST BEs from Zn (2p)
NIST Database defines Adventitious Hydrocarbon C (1s) BE = 284.8 eV for all insulators.

Histogram indicates Zn (2p) BE = 49.6 eV for Magnesium Metal (Zn)
based on 15 literature BEs
Histogram indicates Zn (2p) BE = 50.7 eV for Native Magnesium Oxide on top of Magnesium based on 18 literature BEs



Table #6

NIST Database of Zn (2p) Binding Energies

NIST Standard Reference Database 20, Version 4.1

Data compiled and evaluated
by
Znexander V. Naumkin, Anna Kraut-Vass, Stephen W. Gaarenstroom, and Cedric J. Powell
©2012 copyright by the U.S. Secretary of Commerce on behalf of the United States of America. Znl rights reserved.